00:01
So welcome back to the
ECG lecture series.
00:04
We just heard about heart block and
some heart blocks are treated with pacemakers
and so let’s talk about pacemakers.
00:12
Pacemakers are an amazing development.
00:14
They were developed back in the
1960s and they were developed for,
“Patients whose heart was too good to die.”
What does that mean?
That means people develop complete heart block,
they had heart rates in the 20s and 30s.
00:27
The cardiac output went down,
the kidneys failed and the patients died.
00:31
At autopsy, the heart looked perfect.
00:33
The muscles was good, the valves were good,
the coronary arteries were good -
they died because of an abnormality in the
electrical system that led to the development of pacemakers,
which were quite large at that time,
placed under the skin in the abdomen.
00:47
Nowadays, the average pacemaker’s a little bit
bigger than the watch that you wear on your wrist
and consist most of it consist of the battery,
the lithium battery and a little chip,
that does all of the thinking, is of course very tiny.
01:02
So let’s talk about pacemakers.
01:04
Pacemakers are used to treat
either second degree Mobitz 2 block.
01:10
Remember what second degree Mobitz 2 block is?
That’s where there are drop beats
without lengthening of the PR interval.
01:17
Lengthening of the PR interval can be type 1
secondary AV block - can occur in athletes and in normal people;
but type 2, where you’re suddenly
dropping beats do not occur in normal people,
that means a lot of injury to the conduction system
and you’re very close to third degree AV block
which means complete heart block with a very slow heart rate
and a reduced blood pressure and a reduced heart rate.
01:42
Patients with high grade AV block
are usually symptomatic -
often they faint because
their blood pressure drops.
01:48
They are hypotensive and hypotension with a
very slow heart rate means lowered cardiac output
and if prolonged can result in renal failure and
eventually in death as occurred in the days before pacemakers.
02:03
Syncope or near syncope is very common in these
patients, of course, because of the low blood pressure
and the pacemaker therapy is required to restore
normal heart rate and a normal blood pressure.
02:14
So the pacemakers have a code that tells you
where the pacemakers are being paced
and where they are being monitored and so forth
and I’m gonna take you through the code so
when you see the code somebody says to you,
“Oh, this patient has a DDD pacemaker,”
you’re not standing there thinking,
“I don’t know what that means?”
So, let’s see what that means.
02:37
So, the first letter is where the
active pacing wire is located.
02:42
So, if the first letter is V that means the pacing wire
is in the ventricle and it’s essentially always the right ventricle.
02:49
If there is an A, that means the pacing wire is in
the atrium and that’s almost always the right atrium.
02:56
We usually don’t put pacing wires on the left-side
of the heart because little clots can form there,
break off and go to the brain and cause strokes
so almost all the pacing occurs in the right side -
right atrium or right ventricle.
03:09
And if there’s a D that means there are pacing wires
in both the atrium and the ventricle, D for dual.
03:16
Okay, V means pacing in the ventricle,
A mean pacing in the atrium,
D means pacing in both the atrium and
the ventricle - so that’s the first letter.
03:25
So the example I gave before, DDD,
that means, D, the first letter is D
it means it’s pacing in both the atrium and
the ventricle and I will show you examples of all these as we go along.
03:37
The second letter is where the pacemaker is
sensing the underlying rhythm because, you know,
these days, the pacemakers don’t just keep
firing they have a little chip that reads
whether there’s an underlying rhythm so they
only fire when there’s a failure of the normal beat to occur,
so where do they sense whether
normal beats are occurring?
Well, if the second letter is a V,
they're sensing in the ventricle; if the letter is an A,
they're sensing in the atrium and if the letter is a D,
they’re sensing in both the atrium and the ventricle.
04:10
So remember the example I said, DDD, okay, the first D
says that they are pacing in the atrium and the ventricle.
04:17
The second D says they’re sensing
in both the atrium and the ventricle,
they're looking for P waves and
then they’re looking for QRSs.
04:24
Okay, again, the first two letters are D and D
means the pacemaker is pacing in both the atrium
and the ventricles and sensing in
both the atrium and the ventricles.
04:39
The third letter shows which
heart chamber the pacemaker is inhibited
that is where does it recognize that there’s a regular
beat occurring and therefore it does not pace.
04:50
If it’s inhibited in one ventricle there’s an I and
if it’s inhibited in both the atrium and the ventricle there’s a D.
04:59
There’s another characterization called triggered,
rarely used and really, we almost never see that,
it’s where they're looking to trigger
a beat and have it inhibited,
that really isn't of any major importance,
it’s really, I and D that’s important.
05:14
Let’s see the example. A D inhibited pacemaker
watches first for an atrial depolarization wave
and doesn’t pace if it sees one, then it waits and
it’s looking for a QRS from the ventricle.
05:28
If a QRS occurs, the pacemaker does not pace and if
neither a P wave or a QRS occur, the pacemaker puts both in.
05:36
So for example, it might happen, that there’s a
failure of the atrial depolarization
so the atrium is paced but then there’s
a normal ventricular depolarization.
05:46
The opposite can occur,
there’s a normal P wave but no QRS
so the ventricle is paced or neither occur,
you’ll see both atria and ventricle being paced.
05:57
The fourth letter is really used against
special pacemakers
that can increase the rate when patients
exercise or get excited because
as we talked about before the
pacemakers set at a certain heart rate,
they will always code that heart rate
but what happens if somebody exercises,
they need an increased heart rate so there
are pacers that follow certain physiological variables
for example, skeletal muscle activity or
respiratory rate when they see that increasing,
they know the person’s exercising,
they increased the heart rate
which is what happens in normal people
when they are exercising.
06:35
So when one of those bodily variables,
these are skeletal muscle activity or respiratory rate goes up,
the pacemaker increases the rate of pacing and
the amount that the heart rate will increase is programmed
during implementation but it really
can’t increase above 25 beats;
but again, if you have patients being paced
at 72 and then suddenly, they are exercising,
the other heart rate goes up to 97 so that
increases the cardiac output so it makes their exercise more physiologic.
07:08
The fifth letter is rarely used, it’s mostly for
experimental pacemakers where a multiple site
are being used within the atrium or ventricle and
these are sometimes special arrhythmia pacemakers
for example to control atrial fibrillation,
you almost never see it in routine daily clinical work.
07:25
As noted, many of the pacemakers are
experimental not used routinely
and clinicians almost always deal with a three
to four letter code describing the pacemaker implanted
and we're gonna take some examples.
07:38
So, here is a VVI pacemaker, first V,
where is it sensing? In the ventricle; second V,
where is it pacing? In the ventricle; third letter I,
it’s inhibited when there is a normal QRS.
07:54
So the pacemaker does not pace or is inhibited
when it senses a QRS in the right ventricle.
08:01
What do the four initials stand for
in D-D-D-R?
The first two stand for
sensing and pacing.
08:11
So what does the D stand for?
It stands for dual.
08:15
That means the pacemaker senses
in both the atria and the ventricle.
08:21
And it can pace in either
the atria or the ventricle.
08:25
So, dual sensing, dual pacing.
08:28
The third D
stands for inhibition.
08:32
Again, inhibition is dual.
08:34
So, for example,
if the sinus node is abnormal,
and there's no beat in the atrium,
of course,
the atrial lead will fire
and cause an
atrial impulse
to pass down
into the ventricle.
08:50
However, if there is a
normal sinus node beat
that passes down through the atria,
that pacing is inhibited.
08:59
And the same thing happens
in the ventricle.
09:01
If no beat occurs in the ventricle
after there is an impulse from above
from the atria,
then the pacemaker
supplies that stimulus
and causes a
ventricular depolarization.
09:13
But if there is already a
ventricular depolarization,
the pacing is inhibited
in the ventricle.
09:20
And so you can see the pacemaker
could put in a beat in the atrium,
could put a beat
in the ventricle,
could put beats in both
the atrium and the ventricle,
and be inhibited in both
the atria and the ventricle.
09:32
And then finally, the R.
09:34
This is a pacemaker
that's often put into younger
or very active older people
who want to exercise.
09:41
And what happens is
the pacemaker speeds up
when the patient is active.
09:46
It reads increased respiration,
or increased muscle activity,
and it increases
the heart rate,
thereby increasing the
cardiac output for exercise.
09:56
So that's a clarification of the
D, D, D, R pacemaker.
10:03
In this slide,
we see a little diagram,
a little cartoon, of
how the pacemaker is inserted?
And where it's inserted?
As you know,
noted from earlier,
it's almost always on
the right side of chambers.
10:15
A wire is usually inserted
through a catheter
that's placed in
one of the central veins,
often the subclavian vein
percutaneously,
that is with a needle.
10:27
And the wire is then positioned,
either if it's a pacemaker
that's in both
the atria and the ventricle,
you have two wires,
one putting the atria,
right atrium,
one put in the ventricle,
and that's what this
little diagram shows,
or shows a wire
in that right atrium
and a wire on
the right ventricle.
10:43
Then a little Novocaine
is put in the skin.
10:46
And then a little incision is made,
and a little pocket is made
under the skin.
10:50
And that's where
the actual pacemaker
a box itself is inserted,
and the wire is hooked up to it,
and the incision is closed.
10:59
And I'll show you a picture of that.
11:00
So, you actually see it.
11:02
You'll notice the pacemakers
not in the left atrium,
it's not in the left ventricle,
it's in the right atrium,
and right ventricle.
11:11
And as I said before,
it's because little clots can form
on the pacemaker,
and we don't want them breaking off,
and going to the brain,
and causing strokes.
11:19
And here's you see
the pulse generator,
that's the actual box itself.
11:23
And you see
the two pacing leads:
one in the right atrium,
and one in the right ventricle.
11:29
And so, that's where the wires are,
that are connected to the box
that results in the
generation of the impulses
when the heart itself
does not create an impulse.
11:42
And you can see here,
what I said before,
the pacemaker is inserted
under the skin
with the electrodes
in the right atrium
and the right ventricle.
11:50
And here you can see
the tip of the pacemaker
in the apex
of the right ventricle.
11:56
And here's a chest X-ray
showing a patient
with a VVI pacemaker.
12:01
That is a single lead
in the ventricle.
12:04
So, it's sensing and pacing
in the ventricle,
and it's inhibited
when there's a normal QRS.
12:10
Look, first of all,
at the right side of the diagram,
which is the patient's left chest.
12:15
You can see the pacemaker there,
that's under the skin,
and you'll see
a wire coming from it,
all the way down
and into the right ventricle.
12:24
And so this is a typical pattern
of so called
Posterior/anterior or PA chest X-ray
showing a single lead of
VVI pacemaker in this patient.
12:36
And here we see
the pacer and the wire
that's going to go in.
12:39
You see how small the pacer is
in this person's normal size palm.
12:43
It's a little bigger than a
pocket watch or a wrist watch,
but not very much bigger.
12:48
And you'll see the wires
quite small and thin.
12:52
And here's one
that's been implanted.
12:54
You'll notice
the green arrow,
there a little clips there
closing the incision
that was used
to make the pocket.
13:00
And you'll see a little bulge
just below those clips
that bulges is the pacemaker,
and you see where
there's a little patch there,
a little band aid,
that's where the wire
went in through a jugular
correction, through a
subclavian vein insertion.
13:19
Sometimes,
we'll use the jugular vein
but usually, rarely,
it's usually the subclavian vein.